Temperature rise is the difference between specimen temperature and a defined ambient reference, not simply the hottest colour in a thermal image. Load profile, mounting, cooling, measurement points and stabilization criteria must all be fixed for selection or acceptance data.
DRAWING & SELECTION
Align load, measurement points and ambient reference in time
Record surface temperature Tₛ and the corresponding ambient Tₐ for each point, then calculate temperature rise ΔT at the same time. Freeze the load, mounting and cooling boundaries before comparing conductor, joint and nearby insulation. The diagram shows a DC circuit; define AC or pulsed conditions separately.
| Validation item | Fix before testing | Retain in the report |
|---|---|---|
| Specimen and mounting | Drawing revisions, material/finish, joints, PCB/cables, support and assembly condition | Specimen IDs, mounting photos, fastening procedure and differences from the intended equipment |
| Load and cooling | Continuous/peak current, waveform and duration, orientation, airflow or coolant conditions | Measured current history, fan/cooling settings, adjacent heat sources and deviations from the plan |
| Points and ambient | Point locations, thermocouple type/attachment and ambient sensors positioned by the applicable method | Sensor IDs, calibration, location photos and synchronized Tₛ, Tₐ and ΔT; avoid direct specimen heating of the reference |
| Stabilization and stopping | Applicable standard or approved-plan criterion, observation window, sample interval and stop conditions | Full heating history, when the criterion was or was not met and any early stop; elapsed run time alone is not a criterion |
| Electrical and retest | Voltage-drop points, test current and before/after inspection methods | Load-aligned interface drop, appearance and resistance; perform de-energized checks under the approved procedure |
Check absolute temperature and temperature rise against their respective limits. Shiny-metal images depend on emissivity and reflections; a thermal hot spot cannot replace specified measurement points. Review how sensor attachment or surface treatment changes measurement and heat flow.
Keep the configuration and point histories together
The blank validation sheet covers assembly, calibration, load/cooling, measurement points, ambient reference, stabilization and acceptance references. Use the companion point log for time-aligned Tₛ, Tₐ and ΔT, with original curve files. Neither template supplies a universal limit or replaces report approval.
Download blank temperature-rise validation CSV ↓Use the application to define load and thermal boundaries
Record continuous current, pulses or ripple, energization time, ambient temperature, airflow or coolant, neighbouring heat sources and mounting orientation. Cabinet bars, board-level SMT busbars and liquid-cooled terminals should not share a context-free empirical rise value.
Thermocouples, thermal images and voltage drop answer different questions
Thermocouples provide fixed-point histories. Thermal imaging reveals spatial hot spots but shiny metal can reflect its surroundings because of low emissivity. Four-wire drop helps identify local contact-resistance change. Align all three against the same load timeline.
- Place repeatable points at interfaces, conductor midpoints, ends and nearby insulation
- Record the ambient reference and the full heating curve before stabilization
- Recheck torque, drop and appearance before and after the run
Reports should let another laboratory reproduce the test
Include part revision, materials and finish, torque, cables or PCB, fixture, sensor attachment, load program, environment, instruments, sample interval and stop rule. IEC 60947-1 provides product-standard context for terminal temperature rise, but the applicable project and equipment requirements still set acceptance limits.
Apply this guide to a project
Use product pages to identify the hardware family and the calculator for early estimates only. Acceptance depends on drawings, operating conditions and project tests.
