High-current ATE failures often begin at a contact face or a poorly defined maintenance boundary rather than in the device under test. Replaceable contact modules, measurable drop points and defined cleaning make results easier to trace.
Layer the power path and signal reference
Identify supply, return, Kelvin sense, shield and protective-earth positions. Review high-current terminals for section, contact area and heat rejection. Give low-level measurements a stable reference and fixture space away from switching noise.
Judge contact life with post-maintenance evidence
Finish, elastic travel, guidance and cleaning method all change the contact state. Validate drop, rise, appearance and alignment after the specified mating count, rather than relying on a first-article continuity result.
- Record rise of terminals and supports at test load and representative takt
- Recheck contact resistance, ground continuity, shield contact and insulation spacing
- Define replaceable modules, cleaning media, inspection interval and retirement criteria
ATE acceptance needs traceable fixture evidence
Provide fixture interface, DUT tolerance, test current and duty cycle, mating cycles, contact material and finish, grounding, thermal limit, maintenance steps, spares and record format. For multi-station fixtures, state allowed station-to-station variation and sampling method.