High-current ATE failures often begin at a contact face or a poorly defined maintenance boundary rather than in the device under test. Replaceable contact modules, measurable drop points and defined cleaning make results easier to trace.

01

Layer the power path and signal reference

Identify supply, return, Kelvin sense, shield and protective-earth positions. Review high-current terminals for section, contact area and heat rejection. Give low-level measurements a stable reference and fixture space away from switching noise.

02

Judge contact life with post-maintenance evidence

Finish, elastic travel, guidance and cleaning method all change the contact state. Validate drop, rise, appearance and alignment after the specified mating count, rather than relying on a first-article continuity result.

  • Record rise of terminals and supports at test load and representative takt
  • Recheck contact resistance, ground continuity, shield contact and insulation spacing
  • Define replaceable modules, cleaning media, inspection interval and retirement criteria
03

ATE acceptance needs traceable fixture evidence

Provide fixture interface, DUT tolerance, test current and duty cycle, mating cycles, contact material and finish, grounding, thermal limit, maintenance steps, spares and record format. For multi-station fixtures, state allowed station-to-station variation and sampling method.

Apply this guide to a project

Use product pages to identify the hardware family and the calculator for early estimates only. Acceptance depends on drawings, operating conditions and project tests.

Sources

  1. IEC 61010-1 safety requirements for measurement, control and laboratory equipment↗
  2. IPC-A-610 acceptability of electronic assemblies↗
  3. NASA electrical bonding and mating-surface preparation↗