Connect a Cu-Al busbar SEM-EDS line scan to its specimen location, original spectra and micrograph. The curve can show local composition changes, but sampling step is not spatial resolution, and transition width is not automatically true reaction-layer thickness. Phase and bonding conclusions need separate evidence.

01

Identify the interface being scanned

ISO 22309:2011 covers quantitative point/area EDS in SEM or electron-probe analysis and emphasizes correct element identification. We recommend documenting composite geometry, section direction and processing state, overlaying scan endpoints on a scaled original image with copper, aluminum and interface locations.

Give each scan an identifier linked to its specimen and batch, distinguishing straight sections, hole edges and bends. Review preparation scratches, smearing and contamination. A curve without a sampling map cannot establish which processing region it represents or qualify the whole interface.

02

Separate step spacing, interaction volume and spectral lines

Oxford Instruments explains that lower accelerating voltage generally reduces interaction volume but must still excite useful lines with adequate signal. Retain voltage, acquisition settings, spacing, line selection and analysis approach. Dense sampling alone does not establish equivalent fine-structure resolution.

Retain original spectra and background-processing information, identifying counts, mass fraction or atomic fraction on the axis. Unlike quantities cannot be treated as one concentration scale. Investigate small changes and unexpected peaks against element identification, contamination and settings; adjacent colors or normalized totals do not prove accuracy.

03

A composition gradient does not identify a phase

Review the observed transition against interaction volume and section condition. Spatial averaging may contribute to the apparent width; do not report it directly as true reaction-layer thickness. For dimensional acceptance, define boundaries and a suitable method, retaining independent thickness results and uncertainty.

Oxford's combined EDS/EBSD approach uses chemistry with crystal-structure information for phase identification. A Cu-Al atomic ratio or gradient alone cannot confirm a particular intermetallic phase. Agree suitable crystallographic evidence when needed. Composition checks also do not replace bond strength, resistance or lifetime review.

  • Link scans, images and specimens
  • Retain quantities and settings
  • Separate composition, phase and performance
04

Frequently asked questions

Does dense sampling resolve an equally fine layer? Review interaction volume, signal and preparation; step spacing is not resolution.

Can matching Cu-Al ratios confirm a phase? Suitable crystal-structure evidence is still needed.

05

RFQ and interface-analysis inputs

Provide section drawings, target processing regions, batch identity and destructively testable specimens. Agree scan locations and original-spectrum delivery. Separate composition, thickness and phase objectives, confirming method suitability before choosing coverage and release criteria.

Apply this guide to a project

Use product pages to identify the hardware family and the calculator for early estimates only. Acceptance depends on drawings, operating conditions and project tests.

Sources

  1. ISO 22309 quantitative EDS scope↗
  2. Oxford EDS interaction volume and resolution↗
  3. Oxford combined EDS/EBSD phase identification↗